1、Six-MonthClinicalandAngiographicResultsofaDedicatedDrug-ElutingStentfortheTreatmentofCoronaryBifurcationNarrowingsa,abcEberhardGrube,MD*,LutzBuellesfeld,MD,FranzJ.Neumann,MD,StefanVerheye,MD,deafAlexandreAbizaid,MD,DougalMcClean,MD,RalfMueller,MD,AlexandraLansky,MD,ffagRoxanaMehran,MD,RicardoC
2、osta,MD,UlrichGerckens,MD,BrettTrauthen,MS,andhPeterJ.Fitzgerald,MDPercutaneousinterventionforcoronarybifurcationlesionshasbeenassociatedwithin-creasedclinicalcomplicationratescomparedwithnonbifurcationlesions,primarilyasaresultofrestenosis.Therefore,thereisaneedfornewtechniques.Thepurposeofth
3、isstudywastoevaluateanewdrug-elutingstentandimplantationtechniqueforthetreatmentofdenovocoronarybifurcationlesions.TheAxxessPlustrialwasaprospectivemulticentersingle-armstudythatenrolled139patients.Eachpatientreceivedaself-expanding,con-icallyshapednickel-titaniumAxxessPlusbiolimusA9-elutingst
4、entatthelevelofthecarina.Dependingonthelesionanatomy,additionalnonstudystentswereplaceddistallyifnecessary.Clinicalandangiographicfollow-upwerescheduledat6monthsaftertheprocedure.Theoverallrateoftargetlesionrevascularizationwas7.5%at6months.Ameanof2.4stentswereimplantedperpatient;51.2%ofpatien
5、tsreceivedastenttothesidebranch,29.4%receivedballoonangioplastyonly,and20.6%ofsidebrancheswerenottreated.In-stentlatelossintheAxxessstentswas0.09mm.Incidencesofangiographicin-stentrestenosiswere7.1%intheparentvesselstentsand9.2%inthegroupreceivingstentsinthesidebranch(7.9%excludingbaremetalste
6、ntsplaceddistaltotheAxxessstent),comparedwith25%forballoonangioplastytreatmentand12%fornotreatment.Latestentthrombosiswasobservedin3cases,2ofwhichwereassociatedwithconfirmedprematurecessationofantiplatelettherapy.Inconclusion,theAxxessPlusconicalstenteffectivelytreatsbifurcationlesionsaloneorin
7、conjunctionwithotherdrug-elutingstents.©2007ElsevierInc.Allrightsreserved.(AmJCardiol2007;99:1691–1697)Inthestentera,percutaneouscoronaryinterventionsinthesidebranchostium.Serialintravascularultrasound(US)bifurcationsareassociatedwithin