EPA METHOD 7062 Antimony and Arsenic (Atomic Absorption, Borohydride Reduction)

EPA METHOD 7062 Antimony and Arsenic (Atomic Absorption, Borohydride Reduction)

ID:44077556

大小:165.28 KB

页数:8页

时间:2019-10-18

EPA METHOD 7062 Antimony and Arsenic (Atomic Absorption, Borohydride Reduction)_第1页
EPA METHOD 7062 Antimony and Arsenic (Atomic Absorption, Borohydride Reduction)_第2页
EPA METHOD 7062 Antimony and Arsenic (Atomic Absorption, Borohydride Reduction)_第3页
EPA METHOD 7062 Antimony and Arsenic (Atomic Absorption, Borohydride Reduction)_第4页
EPA METHOD 7062 Antimony and Arsenic (Atomic Absorption, Borohydride Reduction)_第5页
资源描述:

《EPA METHOD 7062 Antimony and Arsenic (Atomic Absorption, Borohydride Reduction)》由会员上传分享,免费在线阅读,更多相关内容在行业资料-天天文库

1、METHOD7062ANTIMONYANDARSENIC(ATOMICABSORPTION,BOROHYDRIDEREDUCTION)1.0SCOPEANDAPPLICATION1.1Method7062isanatomicabsorptionprocedurefordetermining1µg/Lto400µg/Lconcentrationsofantimonyandarsenicinwastes,mobilityprocedureextracts,soils,andgroundwater.Method7062isapprovedforsamplematrice

2、sthatcontainuptoatotalof4000mg/Lconcentrationsofcobalt,copper,iron,mercury,ornickel.Asolidsamplecancontainupto40%byweightoftheinterferentsbeforeexceeding4000mg/Linadigestedsample.Allsamplesincludingaqueousmatricesmustbesubjectedtoanappropriatedissolutionsteppriortoanalysis.Spikedsampl

3、esandrelevantstandardreferencematerialsareusedtodeterminetheapplicabilityofthemethodtoagivenwaste.2.0SUMMARYOFMETHOD2.1SamplesarepreparedaccordingtothenitricaciddigestionproceduredescribedinMethod3010foraqueousandextractsamplesandthenitric/peroxide/hydrochloricaciddigestionprocedurede

4、scribedinMethod3050(furnaceAAoption)forsediments,soils,andsludges.Excessperoxideisremovedbyevaporatingsamplestoneardrynessattheendofthedigestionfollowedbydegassingthesamplesuponadditionofurea.L-cysteineisthenaddedasamaskingagent.Next,theantimonyandarsenicinthedigestarereducedtothetriv

5、alentformswithpotassiumiodide.Thetrivalentantimonyandarsenicarethenconvertedtovolatilehydridesusinghydrogenproducedfromthereactionoftheacidifiedsamplewithsodiumborohydrideinacontinuous-flowhydridegenerator.2.2Thevolatilehydridesaresweptinto,anddecomposein,aheatedquartzcelllocatedinthe

6、opticalpathofanatomicabsorptionspectrophotometer.Theresultingabsorptionofthelampradiationisproportionaltothearsenicorantimonyconcentration.2.3Thetypicaldetectionlimitforthismethodis1.0µg/L.3.0INTERFERENCES3.1Veryhigh(>4000mg/L)concentrationsofcobalt,copper,iron,mercury,andnickelcancau

7、seanalyticalinterferencesthroughprecipitationasreducedmetalsandassociatedblockageoftransferlinesandfittings.3.2Tracesofperoxidesleftfollowingthesamplework-upcanresultinanalyticalinterferences.Peroxidesmustberemovedbyevaporatingeachsampletoneardrynessfollowedbyreactionwithureaandallowi

8、ngsuf