lenstar900与接触性a超测量白内障患者眼轴的对比研究

lenstar900与接触性a超测量白内障患者眼轴的对比研究

ID:22623138

大小:74.78 KB

页数:7页

时间:2018-10-30

lenstar900与接触性a超测量白内障患者眼轴的对比研究_第1页
lenstar900与接触性a超测量白内障患者眼轴的对比研究_第2页
lenstar900与接触性a超测量白内障患者眼轴的对比研究_第3页
lenstar900与接触性a超测量白内障患者眼轴的对比研究_第4页
lenstar900与接触性a超测量白内障患者眼轴的对比研究_第5页
资源描述:

《lenstar900与接触性a超测量白内障患者眼轴的对比研究》由会员上传分享,免费在线阅读,更多相关内容在学术论文-天天文库

1、Lenstar900与接触性A超测量白内障患者眼轴的对比研宄【摘要】目的探讨Lenstar900与接触性A超测量白内障患者眼轴的差异,评价Lenstar900测量眼轴的准确性。方法150例白内障手术患者,依据测量方式不同分为Lenstar900测量组与接触性A超测量组,各75例。分别应用Lenstar900与接触性A超两种方法同时由同一医师测量每例患者眼轴,且均由同一医师行白内障超声乳化联合人工晶体植入术,将测得的眼轴计算人工晶体度数作为手术植入人工晶体的度数。术后观察患者角膜情况、前房深度、术后视力,并于术后3个月检查患者术眼屈光情况,对比分析两种测

2、量方法有无差异性。结果接触性A超测量组与Lenstar900测量组测量眼轴结果比较差异无统计学意义(P〉0.05),所有患者顺利完成手术,无并发症,术后部分患者角膜水肿,绝大部分1周内水肿消退,前房深度正常,均较手术前加深,术后视力恢复至0.5〜1.0;术后3个月两组实际屈光范围[(-1.00±0.25)DSVS(-0.50±0.25)DS]及实际与目标屈光绝对差值[(0.46土0.45)DSVS(0.15土0.35)DS]比较差异有统计学意义(P眼轴DOI:10.14163/j.cnki.ll-5547/r.2016.19.015Comparativ

3、eresearchbetweenLenstar900andcontactAultrasoundinaxisoculimeasurementofcataractpatientsYEWei-ming,ZHANGChun-ling,HUANGYong-quan,etal.ZhongshanCityShiqiSuHuazanHospital,Zhongshan528402,China【Abstract】ObjectiveToinvestigatedifferencebetweenLenstar900andcontactAultrasoundinaxisocul

4、imeasurementofcataractpatients,andtoevaluateaccuracybyLenstar900inaxisoculimeasurement.MethodsAtotalof150cataractpatientsweredividedbydifferentmeasurementwayintoLenstar900measurementgroupandcontactAultrasoundmeasurementgroup,with75casesineachgroup.Thetwogroupsreceivedrespectivel

5、yLenstar900andcontactAultrasoundinaxisoculimeasurementbythesamephysician,whoalsoaccomplishedtheirphacoemulsificationandintraocularlensimplantation.Dataofaxisoculiwastakenincalculationforintraocularlensdegree.Postoperativeobservationwasmadeoncorneacondition,chamberdepth,postopera

6、tivevision,andrefractionexaminationafter3monthsforcomparisonbetweenthetwomeasures.ResultsTherewasnostatisticallysignificantdifferenceofaxisoculibetweenLenstar900measurementgroupandcontactAultrasoundmeasurementgroup(P〉0.05).Allpatientsreceivedsuccessfuloperationwithoutcomplicatio

7、ns.Cornealedemaoccurredinsomepatients,andmostlyrelievedafter1week.Patientshadnormalchamberdepth,whichwashigherthanthatbeforeoperation.Postoperativevisionrecoveredto0.5〜1.0.Thedifferenceofactualrefractiverange[(-1.00+0.25)DSVS(-0.50±0.25)DS]andabsolutedifferencebetwenactualandtar

8、getrefraction[(0.46土0.45)DSVS(0.15±0.35)DS]hads